Wilmad NMR tubes are not 'analytically clean' when delivered to you.
So if your NMR samples require scrupulously clean glass, follow the procedures
below for Difficult Cleaning Problems to assure your sample purity
is never jeopardized. Since NMR tubes are formed over a metal mandrel and
certain organic lubricants are used, these cleaning steps will assure that
any trace organic or inorganic residues from these procedures is removed.
When you invest in high quality precision NMR Sample Tubes, you expect high
resolution and sensitivity. Proper cleaning procedures can help you preserve
the quality of your investment. Since the purpose of an NMR Sample Tube
is to confine a liquid sample in a perfectly cylindrical volume within the
spectrometer probe, the degree to which the tube accomplishes this determines
the quality of the sample tube. Improper cleaning can damage NMR tubes and
reduce your apparent spectrometer performance.
You should never use a brush or other abrasive materials to clean NMR tubes.
Scratches on the inside surface of the tube allow a portion of the sample
to extend beyond the perfect cylinder defined by the NMR tube. Because the
portion of your sample which fills a scratch on the inner surface of a tube
experiences a different magnetic field than the rest of the sample, lines
will broaden and resolution will deteriorate when you use scratched tubes.
And you'll see a reduction in apparent spectrometer performance, unless
you reshim your spectrometer for each sample. That's a tedious procedure
your investment in high quality tubes was designed to eliminate to begin
with.
Proper cleaning of NMR tubes can be easy or difficult, depending on your
sample. We'll start with simple cleaning situations and move to the harder
cleaning problems. Because even difficult cleaning procedures end with a
proper rinsing, explained under Simples Cleaning of NMR Tubes, you should
be familiar with both cleaning procedures.
Simple Cleaning of NMR Tubes
When cleaning your NMR tubes is as simple as rinsing the tube with water
or an organic solvent, you can rinse them one at a time. Your main concerns,
then, are what to do with the rinsate. And, if you're using Acetone, also
preventing dermatitis that results when oils are removed from your skin
by this potent solvent.
If you rinse a lot of tubes, there are apparatuses available that will make
your job much simpler. Tube washers, listed in the Wilmad NMR Catalog as
Solvent Jet Cleaners, provide an easy way to clean either one or five tubes
at a time. Using a vacuum flask and aspirator, solvent recovery is simple.
And your hands won't be so easily dried out by solvents, either.
A final rinse with Acetone is frequently used to remove the last organic
contents from the tube. When your sample is to be dissolved in water or
D2O, a final rinse with distilled water is usually adequate. You may want
to take steps to remove traces of water from the surface of the tube. Follow
the procedures for deuterium exchange, below.
Difficult Cleaning Problems
Tubes left with samples in them for a period of time frequently present
a more challenging cleaning problem. Sample degradation or precipitation
can cause material to adhere to the inner walls of the tube. Rinsing the
tube doesn't always remove this adhered material. So Wilmad recommends using
strong mineral acids such a concentrated or, in severe cases, fuming Nitric
Acid soaks of 1-3 days, as needed. Nitric Acid can oxidize many organic
chemicals and dissolves most inorganic materials, as well. Wilmad doesn't
recommend using Chromic Acid, since residual Chromium can often adversely
affect NMR experiments. Chromic Acid, while a stronger oxidizer, can leave
paramagnetic Chromium VI behind, which can be removed only with repeated
soaks with Nitric Acid. Copious rinsing of NMR tubes washed in acids is
required to assure removal of residual acids. A final rinse with distilled
water or Acetone is also appropriate.
Tubes which contained polymeric samples can be even more difficult to clean.
When the polymers are natural products, like proteins and polysaccharides,
strong acid soaks will usually be sufficient. However, when dealing with
synthetic polymers, the challenge is more severe, since many polymers are
inert to acids or insoluble in organic solvents by design.
Although polymers may not readily dissolve in solvents, it may be possible
to soften them by soaking the tubes in a solvent that swells the polymer.
Then a pipe cleaner might be sufficient to remove the softened material.
It may take some experimentation to find the solvent combination that works
best with your polymer system.
Agitation in an Ultrasonic bath with an appropriate solvent can also help
dislodge stubborn sample residues. However, you should take precautions
to assure that NMR tubes don't touch, since contact and vibrations can fracture
delicate thin wall tubes. Wilmad offers a special tube rack for use in its
Ultrasonic bath that prevents such destructive contact between tubes.
Removing Water from NMR Tubes
Drying tubes at elevated temperatures can reshape and ruin precision
NMR tubes. If you dry tubes in an oven, Wilmad recommends placing tubes
on a perfectly flat tray at 125° C for only 30-45 minutes. Better is
the use of a vacuum oven that will remove water at lower temperatures. In
a flat position, tubes that do reshape could be out-of-round and may not
fit the spinner turbine as well. But they'll not affect the spectrometer
probe adversely. Tubes placed in an oven in a beaker, flask, or tube rack
can bend, increasing Camber (lack of straightness)1. Bent tubes
may still fit the spinner turbine, but can damage or break the NMR probe
insert, a costly repair with many probes.
But even drying at high temperatures doesn't remove water chemisorbed to
the surface of the tube. Thus, the preferred method of water removal is
chemical, not physical, treatment. In most cases, it is the protic content
of water that must be avoided. So Wilmad recom-mends exchanging the protons
of chemisorbed water with a deuterated solvent such as D2O prior to a short
drying period in the oven. A bottle of D2O that isn't being used any longer
is perfect for this purpose.
When water chemically degrades your samples, then removal of water is essential.
Here, reaction of the water with a hydride solution can be used, with caution.
After rinsing the hydride solution, a final rinse with very dry Acetone
can be used to remove rinse solvent prior to oven drying. Cap tubes promptly
to avoid absorption of moisture when removing dry tubes from the oven.
1 See Resonance Report NMR-001
for an explanation of NMR tube specifications, like camber.
|