  NMR-007: Coaxial Inserts in NMR Studies
 |
 |
 |
 |
|
| | |
|
| | | Introduction |
| | |
NMR experiments that involve external reference/lock or that compare
two solutions require two well defined sample volumes within the probe.
Wilmad makes a variety of unique sampling systems that create such distinct
sample volumes. This report will help you select the right insert or coaxial
system for experiments where two volumes are needed. Refer to the table
on the last page for a summary of the inserts Wilmad makes and what experiments
each can be used for.
Challenges in Coaxial NMR Sampling
In superconducting NMR spectrometers, cylindrical geometries work best
in most applications requiring inserts. While spherical sample devices like
Wilmad's 529-A series can be used for some comparisons, it can be difficult
to maintain magnetic field homogeneity over the sample volume. It's easier
to make highly precise cylindrical shapes in Pyrex than to blow precise
spherical shapes, except for small sizes (spherical bulbs <5mm OD). Consider
saving spherical sample inserts for microsampling applications. When determining
magnetic susceptibilities, modulation sidebands caused by imperfect sample
tubes can make precise measurements of chemical shifts impossible. Only
highly skilled manufacturers can offer precision inserts of this type, so
the number of suppliers of coaxial inserts is severely limited.
|
|
| |
Generally, coaxial systems made only of Pyrex will provide a higher degree
of precision than systems incorporating Teflon components. Because Pyrex
is rigid while Teflon is soft and cold-flows, Pyrex can be made to tighter
tolerances than Teflon (see precautions in Resonance Report NMR-010 about
preserving NMR tube precision). If you're performing Variable Temperature
(VT) experiments, all-Pyrex coaxial systems are preferred. With these systems,
components have the same coefficient of expansion and you'll avoid breakage
that can sometimes occur when you mix plastic and Pyrex under VT conditions.
Inserts in External Referencing
External referencing has been replaced in recent years by assignment
of chemical shifts calibrated to the deuterium signal of the spectrometer's
lock channel. But when you have a sample that you shouldn't dissolve in
a deuterated solvent, you'll need an external lock and a reference (internal
is pre-ferred). However, when your samples are too precious to mix with
a reference chemical that is difficult to remove or reacts with your sample,
then external referencing can be employed.
|
|
| |
In these rare circumstances, the external reference must be present only
in sufficient quantity to produce a discernable signal in your spectrum.
Generally, small amounts of the reference are needed if the signal produced
by the reference lies away from the critical regions of the NMR spectrum
of your sample. Wilmad Product Number 516 is a coaxial system that provides
a tiny 50µ space between the insert and outer tube, into which a few
drops of concentrated reference can be placed. Sets are available in sizes
from 3 to 15mm OD. Filling factor is reduced with this system, so use it
only when you aren't sample or solubility limited. If you need a larger
reference volume, e.g. to combine an external reference with an external
lock, follow the recommendations for external locking, below.
The 529 microsample system can also be used with samples studied in 5mm
sample tubes. Place or seal a small volume ("16µl) of reference
material into a 529-D capillary and carefully position in a 5mm tube using
the Teflon Holder (529-B). Make certain the capillary is straight in the
Teflon Holder.
The 529-D is ideal for sample or solubility-limited situations, since you
get spectra with only a small reduction in filling factor.
External Locking
The lock channel of an NMR Spectrometer requires a significant signal
to maintain Rf stability. Thus, coaxial systems like the 516 series should
not be used for external locking. The spacing between the walls of these
inserts and outer tubes (50µ) is not sufficient to support a
spectrometer lock, even if 100% deuterated solvents are used.
|
|
| |
The most commonly used coaxial insert, which will give the best lock signal,
is Wilmad's WGS-XBL Series inserts (X = 4, 5, 8, 10, 12, 13, 15, 16, 18,
or 20). These fit standard Wilmad outer tubes snugly, just like the 516,
but each tapers to a smaller diameter stem in the lower 50mm. For the most
common 5mm size, the stem is 2.5mm OD. The capacity of the stem, where reference
material goes, varies in relative capacity from 5%(for large diameter outer
tubes) to 16% of the sample volume. You place your sample in the annular
space. Add just the amount of sample that will fill the annular space below
the taper of the insert. For approxi-mate volumes of the annular space of
the outer tube (see Wilmad's NMR Catalog).
If you seal your lock/reference solution in the insert, you can switch the
insert from one tube to another simply by rinsing and drying the outside
of the insert. If you're performing VT studies of two solutions, then this
is the ideal coaxial system. Special insert sizes can be made if the standard
sizes aren't right for you.
|
|
| |
Wilmad Product Numbers 517-519 can also be used for external referencing
and magnetic susceptibility comparisons, too. Although confined to the 5mm
OD tube size, these coaxial systems offer a way of varying the ratio between
the reference and sample volumes without buying specially prepared products.
And they're ideal for VT experiments, because the components are all made
from Pyrex. It may be necessary to use a wire hook to position or remove
the Pyrex spacers, which fit snugly into the outer 5mm tube. Avoid scratching
the tubes or positioning the spacers in the Rf coil of the spectrometer
probe.
The least costly approach to external referencing is Wilmad's Teflon Sleeve
Adapter Sets. Useful for probes that take tubes between 8 and 20mm OD, these
permit you to combine standard NMR tubes to make coaxial sets. Because of
the machinability of Teflon, which is soft and easily cold-flows, these
sets are of moderate precision. Teflon Adapters should not be used in VT
studies. The lower adapter is easily positioned. It has notches on the outer
edge that permit you to position it with a wire hook. Avoid scratching the
tubes or positioning the spacers within or near the Rf coil of the probe.
Teflon Sleeve Adapters for 5mm tubes are not available. Use Wilmad product numbers
517-519, above, instead.
Measuring Magnetic Susceptibility
|
|
| |
Because of the high degree of precision of WGS-XBL inserts, you'll get
the best resolution in susceptibility comparisons using these coaxial inserts.
The 517-519 Series of inserts also offers the degree of precision and filling-factor
needed for these critical experiments, but WGS-XBL is easier to use. We
recommend that WGS-XBL inserts be used for all magnetic susceptibility studies,
unless there is an overwhelming reason to avoid them. A reference solution
can be sealed into a WGS-XBL insert that's ordered 50mm longer with a constriction.
Inquire about prices.
Pyrex is a trademark of Corning.
Teflon is a trademark of duPont.
| Product Number |
Description |
Sizes |
Experiments |
|---|
| 516 |
Coaxial Insert |
3-15mm OD |
External Reference |
| 517-519 |
Coaxial Insert |
5mm OD' |
Microsamples, External Reference, External Lock, Susceptbility Measurements |
| 529-A Systems |
Spherical Bulbs |
5 - 20mm OD |
Microsamples |
| 529-D Systems |
Capillary Tube |
5mm OD |
Microsamples, External Reference |
| 529-E Systems |
Cylindrical Bulbs |
5 - 20mm |
Microsamples |
| WG-1132-1 to WG-1132-32 |
Teflon Sleeve Adapters |
8 - 20mm OD |
External Reference, External Lock |
| WGS-XBL |
Stem Coaxial Inserts |
4 - 20mm OD |
Microsamples, External Reference, External Lock, Susceptibility Measurements |
|
EPR-101 |
NMR-001 |
NMR-002 |
NMR-003 |
NMR-004 |
NMR-005 |
NMR-006 |
NMR-007 |
NMR-008 |
NMR-009 |
NMR-010 |
NMR-011 |
NMR-012
Search
Frames | Text Only
To order products, call 1-800-220-5171 in the U.S. and Canada.
International callers, call 609-697-3000.
Copyright 1998 Wilmad
A Division of Scientific Group
Send any questions or problems related to this service to
wilmad-webmaster@obd.com
Web site design and construction by Oki Business Digital
|