The purpose of an NMR tube is to confine a liquid sample in a perfectly
cylindrical volume in a magnetic field while being spun at 25Hz. Since this
description is based on mechanics, it is not surprising that NMR tubes have
been described by structural parameters for many years. The degree to which
an NMR sample tube approaches the ideal mechanical structure defines the
quality of the NMR tube. But since words like 'Camber' and 'Concentricity'
aren't associated by spectroscopists with familiar NMR performance criteria
like spectrometer 'Sensitivity' or 'Resolution,' it is helpful for those
who use NMR tubes to understand how the structural specifications published
by Wilmad relate to NMR instrument performance. If your prestige as a researcher
depends on the excellence of your NMR results, understanding the connection
between these two terminologies will prove a valuable asset to your research
efforts.
What follows is an explanation about each structural parameter used to describe
Wilmad NMR tubes, why each is important, and how each relates to NMR instrument
performance. The explanation will be limited to empirical terms. Wilmad
tubes are grouped into structural Quality Classes and each is given a name,
like Royal Imperial. The minimum tolerances which Wilmad NMR tubes in each
Quality Class must meet are defined in our printed catalogs.
Outside Diameter (OD) and Inside Diameter (ID)
A measure of the distance across the center of the tube from the outer
surface (for OD) and inner surface (for ID), diameters are measured over
a large number of points along the length and around the circumference of
the tube. Best described as a range into which all measurements fall, Wilmad
gauges these dimensions fluimetrically. Meeting stringent standards for
OD and ID assures the sample tube defines a precisely reproducible cylindrical
volume for the sample within the Rf coil. Diameters vary for
Wilmad NMR tubes over a range that is usually <0.001" (0.026mm).
Failure to adhere to strict diameter tolerances can produce a diversity
of undesirable performance characteristics in NMR spectroscopy. If the ID
is too large, vortex plugs and coaxial inserts can move during spectral
accumulation. If the ID is too small, finned vortex plugs might actually
burst the tube, making sample loss, possibly inside the probe, a catastrophic
possibility. When the OD is too small, the tube might slip in the spinner
turbine. This can cause spinning wobble, a major source of modulation sidebands.
Or the sample might shift in the probe or fall completely through the probe
when pneumatically lowered into the magnet. In combination with other imperfections
in an NMR tube, a tube with an OD that is too large could make contact with
the probe insert, damaging the probe, a costly repair. When the failure
to meet OD or ID tolerances is localized in a small area of the tube, portions
of the sample might fall outside the 'perfect' cylindrical volume and be
exposed to a portion of the magnetic field which may not be homogeneous
with the remaining portions of the sample within the 'perfect' cylinder.
You'll know this is the case if you must perform extensive shimming from
one sample to another. In some cases, you may not be able to achieve satisfactory
resolution with further adjustments in magnetic field homogeneity.
Concentricity
A measure of the lack of wall uniformity, concentricity can be thought
of as the degree to which the cylinders defined by the inner and outer surfaces
of the tube are parallel and overlap. Determined using an Indicating Gauge,
this tolerance is reported as a deviation (±) or Total Indicator
Reading (TIR or the absolute variation in wall thickness) as measured at
the ends of a tube during rotation. Concentricity is zero for a 'perfect'
tube. But tubes which conform to TIR tolerances between 0.006" (0.15mm)
and 0.0005" (0.013mm) will spin reliably and provide increasingly
improved spectral resolution. Failure to conform to concentricity tolerances
causes portions of the sample to fall outside the 'perfect' cylindrical
volume, exposing that portion to a magnetic field which may not be homogeneous
with portions of the sample within the 'perfect' cylinder. This leads to
modulation sidebands, which can quickly become significant.
Camber
A measure of the lack of straightness of a tube, camber is determined
with indicating gauges by measuring the deflection at the middle of the
tube held at the ends and rotated. Recorded as a deviation (±) or
Total Indicator Reading (TIR or the absolute value of the deflection range),
tubes with deviations <0.0021" (0.053mm) in Camber can be expected
to spin reliably, barring deviations in other specifications. Higher frequency
NMR spectrometers have placed greater importance on the Camber of tubes
in recent years and stricter tolerances have been required to meet the needs
of spectroscopists fortunate enough to be using the most sophisticated NMR
spectrometers. Modulation sidebands are prominently seen when tubes with
poor Camber are used. When camber becomes marked, contact of the spinning
tube with the probe insert is likely. Instances of probe damage have been
noted when 'cheap' NMR tubes were used. Since only 300µ separates
the tube and insert in the most commonly used probes, meeting strict Camber
tolerances is the most important criterion in maintaining the integrity
of the probes of an NMR Spectrometer and avoiding costly and time-consuming
probe repairs.
The Bottom Line
Of course, a tube must meet its structural specifications to deliver
the performance promised in its description. Unfortunately, not all NMR
tubes available today do. However, each Wilmad NMR tube is inspected by
employees who measure NMR tube parameters daily. This allows Wilmad to provide
NMR tubes of reproducible quality, unlike other manufacturers who make NMR
tubes only occasionally. Unfortunately, bargain-hunting has led to some
NMR disasters. Since the Rf coil in some spectrometers is mounted
onto the probe insert (a thin piece of glass tubing or a cylindrical array
of glass capillaries), the broken inserts resulting from using 'cheap' tubes
cost in excess of $2000 to repair, plus spectrometer downtime. Wilmad NMR
tubes have been supplied with all commercial NMR spectrometers delivered
in the last 25 years. Because they know that only Wilmad NMR tubes reliably
deliver the quality needed to get the optimum spectrometer performance for
which the instrument was designed, every NMR spectrometer manufacturer has
standardized on Wilmad NMR Tubes. Shouldn't you?
|